Accelerator mass spectrometry (AMS) for beryllium-7 measurements in smallest rainwater samples

verfasst von
Collin Tiessen, Daniel Bemmerer, Georg Rugel, Rebecca Querfeld, Andreas Scharf, Georg Steinhauser, Silke Merchel
Abstract

Beryllium-7, mainly measured via γ-spectrometry, is used as a (natural) radiotracer for education and science. For activities < 0.1 Bq and samples containing also longer-lived 10 Be, accelerator mass spectrometry (AMS) is the method-of-choice. We demonstrate that 7 Be and 10 Be can be quantified at the Dresden AMS facility on the same prepared BeO. Detection limits ( 7 Be) are ~ 0.6 mBq. Samples as small as tens of millilitres of rainwater can be chemically processed (after acidification) within a few hours without expensive and slow ion exchange. Isobar ( 7 Li) suppression by chemistry and AMS is sufficient to guarantee for an ultrasensitive, cheap, and fast detection method for 7 Be allowing high sample throughput.

Organisationseinheit(en)
Institut für Radioökologie und Strahlenschutz
Externe Organisation(en)
Helmholtz-Zentrum Dresden-Rossendorf (HZDR)
University of Ottawa
Typ
Artikel
Journal
Journal of Radioanalytical and Nuclear Chemistry
Band
319
Seiten
965-973
Anzahl der Seiten
9
ISSN
0236-5731
Publikationsdatum
15.03.2019
Publikationsstatus
Veröffentlicht
Peer-reviewed
Ja
ASJC Scopus Sachgebiete
Analytische Chemie, Kernenergie und Kernkraftwerkstechnik, Radiologie, Nuklearmedizin und Bildgebung, Umweltverschmutzung, Spektroskopie, Öffentliche Gesundheit, Umwelt- und Arbeitsmedizin, Gesundheit, Toxikologie und Mutagenese
Ziele für nachhaltige Entwicklung
SDG 3 – Gute Gesundheit und Wohlergehen
Elektronische Version(en)
https://doi.org/10.1007/s10967-018-6371-6 (Zugang: Geschlossen)